MP20 Flow Cell Meter

 


PURGE STABILIZATION FLOW CELLS


MP20 MicroPurge Flow Cell


Measures purge water indicator parameters in-line; compact, more portable size (50% smaller than previous models).


Purge Stabilization Flow Cell Specifications E-mail
Model No. MP20 (standard)
MP20D (w/ real time clock / data download)
MP20DT (w/ real time clock / data download / turbidity)
(U.S. Patent Number 6,415,659)
Dimensions 18.5 in. x 15 in. x 6.5 in. (47 x 38 x 17 cm)
Weight 14 lbs. (6.4 kg)
Storage 100 data points (200 opt.)
Stabilization Purge Scan™ technology (U.S. Patent Number 6,415,659)
Case Material structural resin
Keypad 5 keys

Meter Specifications:
Display 3.5 in. (9 cm)
Weight 2.1 lbs (1 kg)
Memory 100 data frames (200 opt.)
Rating Waterproof NEMA 6 [IP67]
Power 3 "C" batteries
Battery Life 12 hours
Temperature 23-122°F (-5 - 50°C)
Cable 6 ft. (1.83 m)

Flow Cell Specifications:
Volume 175 mL
Material rigid urethane
Fitting Type soft-tube "clamp-free"
Fitting Size(s) Inlet: 1/4 in. I.D. x 3/8 in. O.D.
Outlet: 3/8 in. I.D. x 1/2 in. O.D.
Venting Modes horizontal and vertical
Sonde Connection bayonet-style twist mount

Sonde Specifications:
Size 3 in. x 9 in. (8 x 23 cm)
Weight 1.3 lbs. (0.6 kg)

Typical Sensor Performance Specifications:
  Range Accuracy Resolution
Temperature -5 - 50°C (23 - 122°F) ± 0.2°C (0.36°F) 0.01°C (0.018°F
DO 0 to 20 mg/l ± 0.2 mg/l 0.01 mg/l
  0 to 100 mS/cm ±1% of reading ± 1 count 4 Digits
pH 2 to 12 units ± 0.2 units 0.01 units
ORP -999 to 999 mV ± 20 mV 1 mV
Turbidity
(MP20DT Only)
0 to 1,000 NTU ± 5% of reading ± 1 NTU 1 NTU
Salinity* 0 to 70 PSS ± 1% of reading ± 1count 0.01 PSS
*Calculated      

Purge Scan Specifications:
Parameter Stabilization range criteria* pH ± .2 units
DO ± 0.2 mg/l
Conductivity ± 0.020 mS/cm
ORP ± 20 millivolts
Turbidity ± 1 NTU
*NOTE: These are default ranges; settings are adjustable to meet site specifications.
Stabilization basis 3 consecutive readings of selected parameters (one or more of above 5) within above limits, at time interval selected, from 1 to 9 minutes. For example, if 2 minutes is selected, then stabilization would be signaled when 3 consecutive 2-minute intervals showed in-range readings at the end of each interval, requiring 6 minutes.


Elapsed time since Purge Scan initiated shows at the bottom of the screen.

Full data sets are stored at time 0, every 5 minutes, and the 3 consecutive readings which satisfy the stabilization criteria.

Stabilization of purge water quality is signaled visually and audibly when 3 consecutive readings of selected quality parameters (one or more of above 5) fall within the selected ranges. The time interval between readings is selected from 1 to 9 minutes. For example, if 2 minutes is selected as the interval between stabilization readings , then stabilization would be signaled when 3 consecutive 2-minute intervals showed in-range readings at the end of each interval, requiring 6 total minutes of stabilized purge water quality. The elapsed time since Purge Scan was initiated is displayed at the bottom of the screen.

Full data sets are stored at time 0, every 5 minutes, and the 3 consecutive readings which satisfy the stabilization criteria.